| Wavelength Range: |
165-900 nm. Optional to 1200 nm |
| Stray Light: |
Intensity ratio less than 0.001 at 200 nm, less than 0.1 over operating limits. |
| Slit Program: |
Constant bandwidth from 0.01 to 12 nm. Automatic slit closure
option to protect sample and optics from UV light. |
| Baseline Drift: |
Less than +/- 0.15 mdeg per hour drift after 30-minute warm-up. |
| Light Source:
|
450 watt suprasil Xenon lamp controlled by a high stability,
constant current, DC power supply. |
| RMS Noise less than: |
0.1 mdeg at 185 nm
0.8 mdeg at 200 nm
0.06 mdeg at 500 nm |
| Experimental Modes Wavelength:
|
Records CD versus Wavelength at user selected bandwidth and
averaging time.
Data cannot be skewed by selecting improper scan speed. |
| Temperature:
|
Software control of thermoelectric device to maintain constant
sample temperature, generate rising or falling temperature steps.
Records CD versus temperature at fixed wavelength. |
| Kinetics:
|
Measures changes in the CD signal as a function of time.
Measures signal with time constants of seconds and longer. |
| Textbook:
|
"Circular Dichroism and the Conformational Analysis of
Biomolecules." Edited by Dr. Gerald Fasman, Plenum Press. |
| Single Cuvette:
|
Range -10 to +110°c
resolution +/- 0.1
stability +/- 0.3°c |
| Thermoelectric Cuvette Holder:
|
Single Cuvette |
| Thermoelectric Cuvette
Holder:
|
5 cm pathlength |
| Chiller:
|
Heat exchanger for thermoelectric cell holder and 450W Xenon lamp.
950W cooling capacity, 11 lpm, 60 psi. Includes external pressure reducer.
Includes necessary fittings and hoses for system integration. |